Falex 400C Operations Manual Ver. 1.1 (JFTOT)
Falex 400C Operations Manual Ver. 1.1 (JFTOT)
Falex 400C Operations Manual Ver. 1.1 (JFTOT)
Thermal Oxidation
Test Machine
Operation and Maintenance Manual
Falex Corporation 1020 Airpark Drive Sugar Grove, IL 60554 630.556.3669 www.falex.com
VER 1.1 (1-6-12)
Page
Section 2 Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
2.1 Initial Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
2.2 Sample Area and Fuel Preparation . . . . . . . . . . . . . . . . . . . . . . . . . 15
2.3 Instrument and Test Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Section 6 Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
6.1 Charcoal Filter Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
6.2 Desiccant Cartridge Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . 34
6.3 Pump Maintenance (Prime, Purge) . . . . . . . . . . . . . . . . . . . . . . . . . 36
Section 7 Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
7.1 Thermocouple . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
7.2 Differential Pressure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
7.3 System Pressure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
7.4 Factors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
Section 8 Service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
8.1 Software Accessible Service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
8.2 Alarm Log . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
8.3 Fault Indicators . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
8.4 Alarm and Warning Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
WARNING
1 Package of
Funnel filter papers 648-400-011
100
Cleaning pan,
1 648-400-008
stainless steel
Ceramic insulator
1 648-400-009
removal tool
1 Tweezers 648-400-010
Thermocouple - fuel
1 650-009-055
sample
650-030-140 (120VAC)
1 Power cord
or 650-030-150 (220VAC)
1 Stylus 650-204-065
1
Tube Kit
p/n 400-560-003
Falex Corporation | 1020 Airpark Drive, Sugar Grove, IL 60554 USA | (630) 556-3669 | f: (630) 556-3679 | www.falex.com
CERTIFICATE OF INSPECTION
Tube Kit p/n 400-560-003
The enclosed test pieces have been inspected and meet
the following requirements:
Material: 316 Stainless Steel
Manufactured in conformance with:
USA: (630) 556-3669 | [email protected] Europe: +32 (0)16 407 965 | [email protected]
Front USB
Slot
DP Calibration
High Alignment
Line
Desiccant
Chamber
Door
Fault Indicators
(Resettable Fuses)
Touch Screen
Interface On/Off Switch
Air Flow Meter
Ring Stand
Assembly
Coolant
Flow Meter
DP+
Calibration
Alignment
Sample Line
Container
Area
Sample
Test Area
Pump Priming
Door
Interface Panel
USB (2)
Ethernet (1)
Circuit Breaker
Power Plug
Receptical
Interface Panel
USB (2)
Ethernet (1)
Circuit Breaker
Power Plug
Receptical
Aeration line
assembly
400-105-002
Sample output
line assembly
400-105-028
Sample input
line assembly
400-105-020
Pre-Filter Bulkhead
Pre-Filter Screws
O-ring Pre-Filter
620-210-001
2. Plug the power cord (120V, 60Hz, 220V, 50/60Hz) 2. Grasp the ring stand assembly and pull from the
into an appropriate outlet. Electrical source housing [Figure 11a]. After the ring stand assembly is
must have a good ground or an isolated power fully extended from the housing, rotate to the right
with ground. to lock into position [Figure 11b].
3. Connect a Printer to one of the USB or Ethernet 3. Place the funnel into the ring stand and place a
port on the rear of the machine. See page 31 for clean filter paper into the funnel [Figure 12].
instructions to load printer driver.
4. Turn the unit on (by the the rocker switch on the
front of the unit) and allow to warm up for 15 Sample
minutes prior to use. Thermocouple
5. The Falex 400 is shipped with a charcoal filter
located in the beaker stand. Remove the shipping
filter by sliding the filter out of the opening on the Sample-In Line
left side of the beaker stand and replace with a
new filter. It is recommended that this filter be
replaced every 2 weeks to 1 month depending
on usage. See section 7.1 for charcoal filter Aeration Line
replacement instructions.
Funnel
Ring Stand
Beaker
Filter/Funnel
funnel until all the fuel has been placed into the
WARNING filter [Figure 13]. Place the now empty beaker in the
sample-out position at the right of the sample
container area.
5. Place an O-ring on the sample-out line. O-rings
are required on all tubing lines except the aeration
Fuels and solvents are flammable and may cause and sample-in lines. Place a clean small O-ring
irritation to the eyes or skin. Wear protective gog- (650-005-002) at the end of each line before
gles, gloves, and an apron; avoid contact with skin, securing the line in place [Figure 14a and 14b]. Inspect
eyes, and clothing. Use in well ventilated areas and the O-rings and replace if damaged or missing.
keep away from heat or flame. Follow all Material
6. Connect the sample-out line to the appropriate
Safety Data Sheet (MSDS), Hazardous Materials
fitting on the front panel [Figure 4].
Identification System (HMIS), ISO 9000, Lab Stan-
dard Operating Procedures (SOP), and related in- 7. Once the fuel has been filtered, remove the filter
structions. Failure to comply may result in person- and dispose of it according to local regulations.
nel injury or death.
8. Remove, clean, and store the funnel.
4. Using the graduations on the side of a clean
9. Rotate the ring stand assembly to the left and
beaker, pour appropriate amount of fuel into the
store into the ring stand housing of test machine
beaker per the method. A minimum of 500 ml
[Figure 11].
may be used. Slowly add the fuel into the filter/
4. Assemble the test section. Insert the heater Notice: Overtightening will break the ceramic
tube into the heater tube housing so that it is bushings and can cause leaks.
approximately in the center of the housing with
Hex Nut
Heater Tube
O-ring
Heater Tube
Housing
Heater Tube
Shoulder
8. Using clean tweezers, install the DP filter supplied DP Filter DP Filter Housing
in the heater tube kit into the DP filter housing.
Ensure the blue stripe is facing out. If the blue
stripe is missing, then that filter has already been
used so do not use it [Figure 17a]. A new DP filter
must be used for each test. Place a new O-ring,
supplied in the heater tube kit, on top of the DP
filter and push into place [Figure 17b].
9. Attach the large end of the fuel-out line to the DP
filter housing [Figure 18]. Finger tighten.
10. Inspect the O-rings on the fuel-out and bypass
lines. Replace if missing or damaged.
11. Remove the right retaining screw of each bus
bar cap using the hex wrench supplied. Loosen Figure 17a DP Filter Installation
the left retaining screw of each bus bar cap
using the hex tool, so that the cap is free to pivot O-ring DP Filter Housing
on the screw [Figure 19]. To remove any fuel that
may have seeped on to the bus bars, clean the
bus bar contacts and bus bar caps, focusing
on the portions that comes in contact with the
heater tube.
Fuel-Out
Line To
DP- Fitting
Fuel Bypass
Line To DP+
Fitting
Figure 21 Tube Aligned With Top of Figure 22 Fuel-Out and Fuel Bypass Lines
Bus Bar
16. Begin assembling the pre-filter by placing the Place the lower housing onto the upper housing.
pre-filter O-ring on the lower housing [Figure 23]. Line up the screw holes on the upper and lower
housings and insert the screws through the lower
17. Place the pre-filter screen into the upper housing housing into the upper housings [Figure 23]. Finger
[Figure 23].
tighten the screws with the hexagon wrench.
18. Using clean tweezers, remove a new pre-filter filter 19. Inspect the O-ring on the end of the pre-filter
paper from the protective case and place onto the assembly. Replace the O-ring if missing or
pre-filter screen. damaged.
Note: There are two blue paper spacers 20. Connect the pre-filter assembly to the pre-filter
between each pre-filter filter paper. Do not bulkhead fitting and finger tighten [Figure 24].
use blue spacer paper. Use only the white
pre-filter element. Use of the blue spacer
paper will cause the test to fail.
Pre-Filter Screws
Lower Pre-Filter
Housing
Pre-Filter O-ring
Pre-Filter
Bulkhead Fitting Pre-Filter Assembly
Pre-Filter Paper Filter
Upper Pre-Filter
Housing
Assembly
21. Attach the pre-filter assembly to the assembled Note: The heater tube thermocouple can be stored in
test section using the fuel input line [Figure 25]. the hole found on the top right of the beaker stand to
Inspect the O-rings on both ends of the sample the left of the bus bar area when not in use [Figure 26b].
line union. Replace if missing or damaged.
24. Check all of the fittings to ensure they are
22. Finger tighten the fittings until snug. finger tight.
23. Plug in the heater tube thermocouple. Carefully 25. Close the acrylic cover to the sample test area.
insert the thermocouple tip into the hole on the top Leaving the cover open will cause an alarm while
of the heater tube and lower until it stops. Rotate a test is running.
the thermocouple retaining clip to the left to lock
the thermocouple into place [Figure 26a and 26b]. If this is the initial setup or if the F400 has not been oper-
ated for 15 days it is recommended that the automated
Purge be run prior to any tests being performed on the
unit. Refer to section 6.3.2.
Assembled
Test Section
Pre-Filter Assembly
Test Section
Heater Tube
Thermocouple Insert Thermocouple Retaining Thermocouple
Retaining Clip into top of Tube Bus Bar Screws Clip in Place Installed
Retaining Clip
It is recommended that the user familarize themself with Lower right corner buttons: Usually there is a back-
the software by reading this section prior to using any ward arrow, forward arrow or both [Figure 29]. The back-
portion of it. ward arrow directs the user to the previous screen and
the forward arrow advances to the next screen. In many
instances the forward arrow will not appear until certain
3.1 Falex 400 Software Program conditions are met on the current screen (such as; man-
General Notes: Various software screens have datory test information entered, a button pushed, a tem-
similar buttons which are discussed below. perature value reached).
Header: Each screen has a blue shaded header. Lower left corner buttons: Buttons in the lower left are
The center of the header has the screen title with the for user assistance [Figure 29]. Various screens during a
date and time displayed underneath. The time is shown running test have a “Status’ button which activates a Sta-
in standard or military format depending on the user tus screen. The Status screen shows live feedback of
preference. The headers right side may have an abort critical test data and where that data is taken from.
button or a home button. The home button returns to
the Main screen while the abort button will end the
current process. The ‘?’ invokes the software version
builds on the main screen [Figure 28]. The '?' invokes a
reference for help on all other screens [Figure 27].
Help Location
in Operation
Aborts test or Manual
procedure
Additional
pages
Main Screen In
Figure 31
Ready State
Figure 30 Additional Page Arrow
on Instruction Pages Temperature
Throughout various screens will be 'factory' buttons.
must be
‘Factory’ buttons are solely for the use of authorized
below 40ºC
factory personnel and access to this area is protected.
before ‘Run
Test’ button
appears
3.2 Run Test
If this is the inital setup or if the unit
has not been operated for 15 days it is
recommended that the automated Purge be
run prior to any tests being performed on
the unit. Refer to section 6.3.2. Figure 32 Main Screen Waiting For Heater
Tube Temperature to Lower
2. Test Type Screen: This screen has the possibility Note: Only tube rating entries of 0, <1, 1,
of eight different test types to choose from <2, 2, <3, 3, <4, and 4 are accepted for the
[Figure 33]. The D3241 ASTM test is always active breakpoint tube reading.
and the button cannot be deactivated. Single,
Breakpoint 1, Breakpoint 2, and the 4 Custom D. Breakpoint 2: Breakpoint 2 runs a matrix of
tests are activated from the ‘Preference’ screen. tests checking both the visual tube reading
value (numeric, abnormal and peacock) and
the maximum ΔP value. All tests run
150 minutes.
Active
NOTE: At the time of this printing Breakpoint 2 is not
included in the test series matrix.
Figure 41a Test run time Figure 42 Test Completion Screen Waiting
For Tub Temperature To Cool
Note: If the Falex 400 has had a power cycle since the
last test, data transfer cannot be performed.
The name of the last test run is on the top line on this
screen. This is the file that the first three buttons refer to
[Figure 45]. The user can view the test data on the touch
screen, transfer the data from the PLC storage onto the
computer, print this report and/or copy files from the
touch screen to a USB device or Ethernet connection.
Complete data records of DP mmHg, heater tube tem-
perature and system pressure are taken every minute
during a test and are stored in the data file. To retrieve
the data file the operator needs to ‘Transfer’ the data file
and then ‘Copy’ the file to an external folder. Data files
are saved in .csv format and can be viewed using ap-
Figure 46 Viewing Data
propriate software. Page 1 - Test Information
Access
the data
Name of
data file
that View,
Transfer
and Print
refer to
2. Transfer: After a test completion this button 4. Copy: ‘Copy’ opens Windows Explorer. Navigate
MUST be selected to move the test data to the to the “Data” folder under the root directory (400
computer. Upon transfer, data files reside in the (C:)). This folder contains all the data files that
“Data” folder under the root directory (F400 (C:)). have been transferred. Close Windows Explorer
An acknowledgement window appears upon when copying is complete. For copying to external
successful data transfer [Figure 48]. Do not proceed devices there are three USB ports and one
until this acknowledgement appears. Ethernet port located on the 400. Two USB ports
and the Ethernet port are on the convenience
Note: The current test data will be lost at outlet on the back panel. One USB port is located
the start of the next test if this ‘Transfer’ on the front of the unit directly above and to the
button is not pressed or when the F400's right of the drierite window.
power is cycled. The computer has limited
hard drive space and it is recommended Note: Do not delete or rename the Data
that the user regularly perform file folder or data transfer will not occur and test
maintenance on the “Data” folder . data information will be lost.
Access
Preferences A. Test Selection: This section is on the left-
hand side of the screen containing seven box-
es with respective titles on their right. Clicking
on these boxes will put an 'X' in the box and
activate the different test buttons available on
the TEST TYPE screen. To deactivate a test
type, click the boxes once more so that the 'X'
disappears from the box.
A ‘Single’ test allows the user to enter all the
test header information and all the test param-
eters. A single test runs one test with these
parameters. They are not saved when the
Figure 50 Accessing machine‘s power is cycled.
Preferences
‘Breakpoint1’ follows the breakpoint method
described inn ASTM D3241. It is based only
2. Preferences Screen: After the password is on the VTR (visual tube reader) integer value
entered, the Preference screen appears [Figure 51]. entered at the completion of each test section.
There are four areas to this screen. First there is
the 'Test Selection' section to the left-hand side of ‘Breakpoint2’ is a matrix method used to
the screen containing seven test types, a ‘Time’ determine the fuel’s breakpoint but takes into
section containing selections for the date/time consideration the VTR readings with the A
format, an ‘Install Printer’ button for access to (abnormal) and P (peacock) attributes and the
the Windows Control Panel for adding additional DP value. The only test parameter the user
printers, and a value box labeled ‘ASTM Temp, C:’ may enter is the starting temperature.
to set the stored ASTM test temperature. There are four custom tests the user can
choose. Each ‘Custom’ test allows the user to
enter all the test header information and the
test parameters. Once they are entered they
are stored in the unit even through
power cycles.
B. Time: On the right-hand side of the Prefer-
ences screen, there is a box labeled 'Time.' In
this box, there are two buttons 'Mil' and 'Std.'
These stand for 'Military' and 'Standard' time
respectively and will change the format of the
time display on the machine.
Desiccant may cause irritation to the eyes or skin.
Wear protective goggles, gloves, and an apron;
avoid contact with skin, eyes, and clothing. Use
in well ventilated areas and keep away from heat
or flame. Follow all Material Safety Data Sheet
(MSDS), Hazardous Materials Identification System
(HMIS), ISO 9000, Lab Standard Operating Pro-
cedures (SOP), and related instructions. Failure to
comply may result in personnel injury or death.
Desiccant Housing
Retainer Clips
Desiccant Cartridge Inlet/Outlet Ports
Rubber Bushings
Black Handle
Plastic Syringe
6.3.3 Purge
The Purge cycle should be performed the first time the
Falex 400 is started, whenever the unit has sat unused for
30 days or more or whenever there is air in the lines and
the prime pump sequence failed to remove all the air.
Calibrations should be preformed regularly on the 400 7.1.1 Thermal Calibration Lo Check
in accordance with the method. Refer to each method
for it is required calibration schedule. Calibrations should 1. Prepare a low temperature standard, from an ice
be performed on the heater tube thermocouple, the dif- bath of distilled water and crushed distilled water
ferential pressure transducer and the system pressure. ice (an insulated cup works best).
‘Thermocouple’ allows for a calibration of the temperature 2. Place the heater tube thermocouple in the
control system using ice and the eutectic temperature of ice bath.
lead to calibrate the heater tube thermocouple. The dif- 3. Stir the thermocouple in the bath watching the
ferential pressure is the measurement across the DP test tube temperature on the display [Figure 64].
filter. Excessive differential pressure (DP) is an indication
of poor fuel quality. ‘Diff Press’ provides a calibration of 4. Continue stirring until the tube temperature no
the differential pressure transducer. ‘Sys Press’ allows for longer drops and holds constant for 20–30
adjustment of the pressure measurement against a NIST seconds. Press the ‘Accept’ button. This will
certified reference pressure gage. store the observed low temperature and will
initiate the Thermocouple Hi Check display. The
Select thermocouple reading should be accepted if
Thermocouple the displayed value is 0.1˚C ±2˚C. Should the
thermocouple not reach this value, verify the ice
bath is not contaminated. If it still will not reach an
acceptable value, replace the thermocouple.
Current
Thermocouple
Reading
Figure 63
7.1 Thermocouple
The auto calibration performs a calibration of the heater Figure 64 Thermocouple Lo
tube thermocouple by a ‘low’ calibration check using an Check Screen
ice bath and a ‘high’ calibration procedure using the lead
calibration assembly. It is critical that the following steps
be followed carefully. Failure to have the thermocouple in 7.1.2 Thermocouple Calibration
the correct position during each step can cause the unit
to continue to apply heat to the bus bars or fail to heat the
Hi Check
bus bars. Should a step not complete as documented 1. Install the lead calibration check assembly in the
within seven minutes of initiation, shut the unit off upper/lower bus bars. Make sure the top of the
from the power switch on the front, allow the unit to assembly is flush with the top of the upper bus bar
cool for a minimum of 20 minutes before restoring and that the bus bar screws are tight.
power and re-initiating calibration.
Manual insertion and removal of the
1. [Figure 63] Fromthe Main Menu, go to the thermocouple is required during the Hi
Maintenance > Calibrate > Thermocouple display Check. Care should be taken during all
and select ‘Auto Calib’ button. This will reset the steps as the bus bars are extremely hot
previous calibration factors to the default values and can carry high current. Only insert
(0 ºC factor = 0 and 327.5 ºC factor = 327.5). It and remove the thermocouple by the
will also call up the ‘Thermocouple Calibration Lo insulated leads.
Check’ display.
2. Insert the heater tube thermocouple through
the top bus bar into the assembly so that the
thermocouple is in contact with the lead.
3. Push the ‘Start‘ button to begin the heat cycle 10. If the eutectic temperature capture was not
that will raise the lead assembly temperature successfully detected, a popup message
to 350 °C. Once pushed, the ‘Heat’ button will display a failure and abort the calibration
will disappear. procedure. It is suggested that the test be
repeated to confirm that the thermocouple is
4. The lead temperature will begin to rise. The
defective. Push the ‘OK’ button on the popup
thermocouple must maintain contact with the
display for a one minute heating cycle to begin to
lead. An occasional light pressure on the top of
allow for thermocouple removal. As described in
the thermocouple wires may be required. Once
steps 9 and 10 above. Remove the thermocouple
the temperature is above 327.5 °C, the lead will
and inspect it for damage. Allow the lead
melt down into the fixture cup. Make sure that the
calibration check assembly to cool. Repeat
thermocouple remains in contact with the lead
the calibration.
by pushing it down to the lowest position with the
tip centered in the cup (do not let thermocouple 11. If the thermocouple fails to capture a satisfactory
touch the sides of the lead cup). eutectic temperature detection again, it should
be replaced.
The Following steps are automated
Real-Time
5. Once the temperature is above 340 °C for 1 Temperature
minute to allow the lead temperature to equalize,
the cooling sequence will be initiated so that the
eutectic temperature can be detected.
6. The system will attempt to automatically detect the
eutectic temperature of the lead by lowering the
temperature below 327.5 °C to a value of 310 °C.
When the temperature reaches 310 °C, a popup
display will appear. It will inform the operator if the
eutectic temperature detection was successfully
detected or if it failed. If it was successful, the
temperature detected will be displayed and the
date it was detected and stored will be displayed.
The stored low temperature point from step 5 and Figure 67 Thermocouple Hi Check Screen
the eutectic temperature will be moved to the
thermocouple calibration factors until the next time
an auto calibration is initiated. 7.2 Differential Pressure
7. If the eutectic temperature was successfully 1. Falex 400 must be on for at least 30 minutes
detected, push the ‘OK’ button on the popup before following this procedure.
display. This will cause the popup to disappear.
The temperature will increase to melt the lead for 2. Fill the calibration bottle ¾ full with fuel.
thermocouple removal. A message is displayed
3. Be sure the sample out line is in place and
notifying the operator to remove the thermocouple
empties into a beaker. The aeration and sample in
from the lead once the temperature reaches
lines can remain in place or they can be closed by
>330 °C. A one minute timer is started to allow
attaching nuts to the bulkhead fittings.
thermocouple removal.
4. The heater tube and prefilter assemblies are not
8. Once the one minute timer has completed or
used for this calibration and need to be removed.
the thermocouple has been removed, heating
Place a nut over the bulkhead where the prefilter
is stopped. The operator must remove the
assembly attaches to the unit. Cap the DP+
thermocouple from the lead at this time and
bulkhead but leave the DP- port open.
clean the thermocouple with a cloth to remove
any residual lead on the thermocouple. Dispose 5. Get to the DP Calibration screen from the Main
of the cloth properly. Wear appropriate gloves screen by the following sequence of buttons:
when handling the hot thermocouple with possible Maintenance-Calibration-DP.
lead residue
6. Upon entering the DP Calibration screen press
9. A four minute timer is initiated to allow for the lead START.
cup assembly to cool. Once the four minutes has
been completed the cooling is stopped..
7. Open the Bypass valve by pushing the Bypass 27. Press the Done button on the screen.
Valve button on the screen until it appears green.
28. Return to the Main screen.
8. Close the System Vent by pushing the System
29. Attach nuts to the DP+ and DP- bulkheads or
Vent button on the screen until it appears red.
reattach the heater tube assembly. Leaving the
9. Attach the calibration bottle to the DP- port on the bulkhead fitting open to atmosphere will cause
Falex 400. fuel to drain from the unit which may cause pump
damage if the unit is run without proper priming
10. Set and leave the bottle on the top of the machine. and purging.
11. Open the DP- Vent on the screen by pressing the
DP- button until it turns green. Live DP
12. Wait until the DP value on the screen holds steady
Reading
(approximately 50 mls have run out of the bottle
and into the outlet beaker).
13. Close the Bypass Valve on the screen by pressing
the Bypass Valve button until it turns red.
14. Close the DP- vent on the screen by pressing the
DP- button until it turns red.
15. Leave the bottle on top of the Falex 400.
16. Remove the nut from the DP+ port and set it aside.
17. Disconnect the bottle from the DP- port and attach
the bottle to the DP+ port. DO NOT CAP THE
DP- PORT. Figure 68 DP Calibration Screen
18. Open the DP+ vent on the screen by pressing the
DP+ button turning it green.
7.3 System Pressure
19. Wait until the DP value on the screen holds steady
(approximately 50 mls have run out of the bottle This calibration is to be factory trained personnel only.
and into the outlet beaker) [Figure 68].
20. Close the DP+ vent by pushing the button on the 7.4 Factors
screen turning it red. “Factors’ displays the most recent values and date for
21. Bring the bottle down from the top of the machine the last calibration of the heater tube thermocouple, dif-
and align the fluid level in the bottle with the center ferential pressure and system pressure [Figure 69].
of the DP+ port alignment line.
22. Hold the DP calibration bottle steady so the fluid
level is level. Hold the bottle in this position for
about 1-5 minutes until the DP value on the graph
remains constant. .
23. Press the Zero button on the screen without
moving the bottle. The Zero offset value on
the screen should now show 0 and show the
today’s date.
24. Raise the fluid level in the bottle to the 400mm
DP calibration high alignment line on the machine
front (near the top right of the Drierite cover).
25. The DP reading on the graph should be 25mmHG
Figure 69 Calibration Factors Screen
+/- 2.5.
26. The fluid bottle may now be set down on the
bench top.
8.1 Software Accessible Service 3. Close DP Pos: Pressing this button manually
operates the DP positive valve. This button is not
Obtaining manual access to machines functionalities is interlocked with the bypass valve and should be
done through the ‘Service’ screen accessed from the used only if the operator has an understanding
maintenance screen [Figure 70]. The user has access to of the system as not to damage the pressure
various controls so operation should only be undertaken transducer.
by someone with understanding of the system.
4. Factory: This is a protected area for the authorized
1. I/O & Vars: [Figure 71] Manual operation of different users only.
controls can be accessed by pressing the large
green or red buttons under Digital Points. The 8.2 Alarm Log
HPLC pump can manual be set to flow at either
3 or 9 ml/min for this service screen. And live Pressing the alarm log button displays the last
[Figure 72]
feedback is seen in the Data section of this screen. 25 warnings and abort alarms.
Description Alarm Abort Condition Lo Limit Hi Limit Alarm Message (HMI) Comments
F10572_F400Manual.indd 43
5 ºC in 10
after pressure achieved.
Heater Tube Tempera- sec af- Heater Tube Failed to
x Possible Cause: Thermocouple
ture Fail To Rise ter press Rise 5 C in 10 s
not installed properly, Thermo-
achieved
couple not in heater tube.
Checked at beginning of test
after pressure achieved. Tem-
90 sec perature must be within range
Section 8: Service
2/3/12 9:13 AM
Description Alarm Abort Condition Lo Limit Hi Limit Alarm Message (HMI) Comments
F10572_F400Manual.indd 44
Aeration aeration. Temperature must be at
High Humidity x time > 30 21% High Humidity high level for 1 sec.
sec Possible Cause: Air pump failure,
humidity sensor failure.
Abort after 1 second. Possible
Heater Tube Thermo- Cause: Excessive deposit on
Heater Tube 'High' Temp x 425 ºC
couple Hi Failure heater tube, Heater tube thermo-
couple failure.
Abort after 1 second.
Possible Cause: Cooling system
Bus Bar 'High' Temp x 80 ºC Bus Bar Temp > 80 C
failure, Thermocouple not in-
stalled properly.
Abort after 1 second.
Possible Cause: Cooling system
Bus Bar 'High' Temp x 60 ºC Cooling System Failure
failure, Thermocouple not in-
stalled properly.
Abort after 1 second.
System 'Low' Pressure x Test active 3.1 Mpa Low System Pressure Possible Cause: Pressure leak,
inlet valve failure.
Abort after 1 second.
System 'High' Pressure x Test active 3.8 Mpa High System Pressure Possible Cause: tubing blockage,
pressure sensor failure.
Temperature must be 'outside' of
≤ 2.5 ºC ≥ 2.5 ºC limits. Abort after 3 seconds.
Heater Tube Temp De- Heater Tube Temp Out
x Test active from set- from set- Possible Cause: Thermocouple
viation of Range
point point failure, excessive buildup on
heater tube.
Abort after 1 second.
Heater Tube Temp Ther- Heater Tube Thermo- Possible Cause: Thermocouple
x Test active -100 ºC
mocouple Failure couple Lo Failure failure, Thermocouple not con-
nected.
Section 8: Service
44
2/3/12 9:13 AM
45
Description Alarm Abort Condition Lo Limit Hi Limit Alarm Message (HMI) Comments
F10572_F400Manual.indd 45
Bus Bar Temp Thermo- Bus Bar Thermocouple Possible Cause: Thermocouple
x Test active -100 ºC
couple Failure Failure failure, Thermocouple not con-
nected.
Abort after 1 second.
Sample Fuel Temp Ther- Sample Thermocouple Possible Cause: Thermocouple
x Test active -100 ºC
mocouple Failure Failure failure, Thermocouple not con-
nected.
Section 8: Service
Warning only.
Safety Cover Missing x Test active Bus Bar Cover Missing Possible Cause: Cover removed
during test, photo eye failure.
Checked after data transfer but-
Data xfer ton selected.
FTP Comm Open Action
Data Transfer Failure x button Possible Cause: Data file not
Failure
selected created, data directory does not
exist, FTP application corrupt.
2/3/12 9:13 AM
Addendum A: Parts Listing 46