General Article
Modification of Suction-Induced Hemolysis During
Cell Salvage
Jonathan H. Waters, MD*
Brandon Williams, BS†
Mark H. Yazer, MD, FRCPC‡§
Marina V. Kameneva, PhD†储
BACKGROUND: The efficiency of red blood cell collection during cell salvage is
dictated by multiple variables, including suction pressure. In this study, we
attempted to determine the influence of suction pressure on the efficiency of cell
salvage and to identify methods for minimizing the impact of suction on salvaged
blood.
METHODS: Whole blood was placed in 60-mL aliquots either in a beaker or on a flat
surface and suctioned at 100 and 300 mm Hg. The amount of hemolysis was
measured and compared under the varying conditions. The experiments were
repeated with the blood diluted with normal saline solution in a 1:1 mix.
RESULTS: Hemolysis ranged from 0.21% to 2.29%. Hemolysis was greatest when
whole blood was suctioned from a flat surface at 300 mm Hg. It was reduced when
the blood was diluted with saline. Blood suctioned from a surgical field during cell
salvage should be done with minimal suction pressures and with the goal of
minimizing blood–air interfaces.
CONCLUSIONS: Significant reduction of blood damage can be obtained by diluting
blood with normal saline while suctioning it from the surgical field. Although
immediate hemolysis due to suctioning was not very high, the red blood cell
damage from suctioning produced by a dynamic blood–air interface might
adversely affect the efficiency of cell salvage.
(Anesth Analg 2007;104:684 –7)
T
here are many benefits of autologous blood conservation, including reduction of demands for allogeneic
blood (1), avoiding the costs of blood products, avoiding the immunosuppressive effects of allogeneic transfusion (2), reduced incidence of transfusion-related
acute lung injury (3), and reduced risk of transfusiontransmitted diseases (4). Several strategies can be
applied to avoid allogeneic transfusion. The primary
techniques involve preoperative erythropoietin administration and iron supplementation, preoperative
autologous donation, acute normovolemic hemodilution, and the application of cell salvage systems. Of
these techniques, cell salvage offers the greatest ability
to avoid transfusion if applied optimally (5).
In an attempt to optimize the capture and return of
red blood cells (RBCs), many cell salvage programs
minimize the suction pressure used to aspirate blood
from the surgical field (6). Minimizing the suction
pressure reduces the mechanical injury to RBCs,
From the *Department of Anesthesiology, Magee Womens Hospital of University of Pittsburgh Medical center; Departments of
†Bioengineering and ‡Pathology, §The Institute for Transfusion
Medicine; and 储McGowan Institute for Regenerative Medicine,
University of Pittsburgh, Pittsburgh, PA.
Accepted for publication November 21, 2006.
Address correspondence and reprint requests to Jonathan H.
Waters, MD, Department of Anesthesiology, Magee Womens Hospital
of University of Pittsburgh Medical Center, 300 Halket St., Suite 3510,
Pittsburgh, PA 15213. Address e-mail to
[email protected].
Copyright © 2007 International Anesthesia Research Society
DOI: 10.1213/01.ane.0000255208.96685.2e
684
which is mostly due to air bubbles mixing with the
blood in the suction cannulae and the tubing connecting the surgical site with the salvage device. The air
aspirated with blood during suctioning produces fastmoving bubbles, which expand and collide in the
negative pressure environment, generating mechanical stress on the RBCs.
Aspiration of air along with blood is a significant
cause of RBC damage during processes other than cell
salvage. Entrained air during cardiotomy suction
damages blood during extracorporeal circulation (7).
The blood-air interface contributes to hemolysis in
artificial organs (8). Air damages blood in bubble
oxygenators as the result of the interfacial denaturation of plasma proteins (9). This causes both sublethal
RBC damage (10) and hemolysis of the RBCs (11,12).
In this study, an in vitro model was developed to
mimic the trauma sustained by RBCs during the
suctioning. We measured the effect of suction tip
immersion in blood when compared with suctioning
from a flat surface to determine ways in which the
mechanical injury might be minimized.
METHODS
On 5 successive days, 11 U of blood group A, citrate
phosphate dextrose preserved and anticoagulated
whole blood, collected from volunteer donors, was
obtained from the blood bank. Informed consent for
use of the blood for research purposes was obtained at
the time of donation, IRB approval was not required.
Vol. 104, No. 3, March 2007
For each experiment, 2–3 U of same type whole
blood were mixed in a 2-L beaker with a plastic
covered metal stirring rod and kept continually mixed
using a magnetic stirrer (Thermolyne, Fisher Scientific, Pittsburgh, PA). For each experiment, the age of
the pooled blood was identical. Before commencing
the experiments, a sample of the blood was taken for
analysis of plasma-free hemoglobin (plfHb) to measure the baseline degree of hemolysis present in the
mixed blood.
Sixty milliliters of blood was taken from the beaker
and placed in a separate small beaker meant to
simulate blood drawn from a deep wound. Similarly,
60 mL of blood was placed onto a flat tray covered
with a textured plastic material. This was intended to
replicate a flat surgical site surface. The blood was
then suctioned from either location using a standard
Yankauer suction catheter tip (Medi-Vac Yankauer
Suction Handle, Catalog K87, Cardinal Health,
McGaw Park, IL). The blood was collected into a
patient suction canister. The tubing connecting the
catheter and a canister was 3 m long with a 6-mm
internal diameter (Medi-Vac nonconductive suction
tube, Cardinal Health). Suction pressure was regulated using the regulated suction from a COBE BRAT
cell salvage device (Arvada, CO). Two suction pressures (100 and 300 mm Hg) were used. The amount of
hemolysis induced by suctioning was calculated for
each pressure. Suction pressure on the machine was
calibrated by the manufacturer during regular machine servicing. Each experiment was performed in
duplicate. In addition, 60 mL of blood was mixed with
60 mL of normal saline and suctioned at 300 mm Hg
from both simulated surgical sites. After each run, the
flow system was thoroughly washed with saline, and
the patient suction canister was replaced.
From the collected blood, a 5 mL sample was placed
into a 7-mL red top vacutainer tube and centrifuged at
3600 rpm for 20 min at room temperature. Blood hematocrit (HCT) was measured (Hematocrit Centrifuge, IEC
Clay Adams) in each sample before centrifugation. The
supernatant from each sample was transferred to a 1.5
mL microcentrifuge tube and recentrifuged at 14,000
rpm for 15 min at room temperature (Eppendorf 5417R
Centrifuge, Hamburg, Germany). The supernatant was
then transferred to a spectrophotometer cuvette (1.5 mL
semimicro-UV methacrylate cuvette, Fisher Scientific,
Pittsburgh, PA) and the plfHb was determined in
mg/dL by measuring light absorbance of the samples at
540 nm (Spectronic Genesys 5 Spectrophotometer, Spectronic Instruments, Columbus, OH). The difference between plfHb in the baseline sample and the suctioned
sample represented an increase in plfHb (hemolysis) due
to exposure to mechanical stress from the suctioning
process.
As the blood samples had different HCT values, and
as mechanical hemolysis depends on HCT, the measured values of plfHb, including those in the baseline
samples, were normalized to the standard HCT value of
Vol. 104, No. 3, March 2007
40%. This normalization was justified by previous findings (13), which showed that, in the range of HCT values
in our samples, there is a linear relationship between
levels of mechanical hemolysis and concentration of
RBCs. This procedure allowed for comparing blood
damage among blood samples with HCT values that
were significantly different (i.e., after dilution of blood
1:1 with saline). Normalization was performed with the
following equation:
Normalized plfHb
ⴝ measured plfHb/(measured HCT/40)
Finally, the percent of hemolysis was calculated using
the formula
% Hemolysis ⴝ [plfHb/wblHb] ⴛ 100
where wblHb ⫽ whole blood total Hb concentration in
mg/dL.
Differences in hemolysis among samples were compared using a one-way analysis of variance with a
Bonferroni’s multiple comparison test. A P value of
⬍0.05 was considered statistically significant. Correlation between the age of the blood and the resulting
hemolysis was performed with Pearsons correlation
test.
RESULTS
Five sets of experiments were performed. All blood
units were type A with ages of 35 days (4 U), 16 days
(2 U), 21 days (3 U), and 24 days (2 U). Hemolysis data
from the pooled experiments under each set of conditions are presented in Table 1. For all of the tested
blood collection conditions the r value correlating the
blood age and change in plfHb was 0.84 (P ⫽ 0.16) for
the 100 mm Hg suction from a surgical surface, 0.76
(P ⫽ 0.24) for the 300 mm Hg section from a surgical
surface, 0.87 (P ⫽ 0.14) for the 100 mm Hg suction
from a beaker, and 0.73 (P ⫽ 0.27) for the 300 mm Hg
from a beaker.
The results are presented after recalculation of
plfHb to a standard HCT level (HCT ⫽ 40%) with the
assumption that hemolysis is linearly proportional to
the concentration of RBCs. As shown in Table 1,
hemolysis was always significantly higher when
blood was aspirated from the flat surface, regardless
of the suction pressure. At a negative suction pressure
of 300 mm Hg, dilution of blood with saline (1:1)
reduced hemolysis by approximately 55% after suction from the beaker and by more than 60% after
collection from the flat surface. Table 2 shows how the
suctioning speeds differed based on suction pressure
and suction site. As would be expected, 300 mm Hg
suction pressure allowed for more rapid suctioning.
However, the site of suction was more important:
suctioning from a bowl was much more rapid than
suctioning from a flat surface. A reduction of blood
viscosity from saline dilution reduced the time of
© 2007 International Anesthesia Research Society
685
Table 1. Relative Hemolysis as a Function of Surface, Suction Pressure, and Dilution
Tests
Increase in plasma-free hemoglobin
from baseline (mg/dL)
Relative hemolysis
(%)
Surgical surface at 100 mm Hg† (n ⫽ 10)
Surgical surface at 300 mm Hg* (n ⫽ 10)
Surgical surface at 300 mm Hg ⫹ dilution (n ⫽ 10)
Beaker at 100 mm Hg (n ⫽ 10)
Beaker at 300 mm Hg (n ⫽ 10)
Beaker at 300 mm Hg ⫹ dilution (n ⫽ 10)
135.5 ⫾ 35.9
304.3 ⫾ 60.6
119.3 ⫾ 15.0
36.0 ⫾ 12.8
63.9 ⫾ 15.2
27.7 ⫾ 8.6
1.02 ⫾ 0.27
2.29 ⫾ 0.46
0.90 ⫾ 0.11
0.27 ⫾ 0.10
0.48 ⫾ 0.11
0.21 ⫾ 0.06
Data presented as Mean ⫾ SEM.
* Surgical surface at 300 mm Hg significantly different than all other groups (P ⬍ 0.001).
† Surgical surface at 100 mm Hg is significantly different from beaker at 100 mm Hg (P ⬍ 0.05)
Table 2. Suction Speed
Time of
suctioning (s)
Testsa
Surgical surface at 100 mm Hg*
Surgical surface at 300 mm Hg†
Surgical surface at 300 mm Hg ⫹
dilution‡
Beaker at 100 mm Hg
Beaker at 300 mm Hg
Beaker at 300 mm Hg ⫹ dilution
75.3 ⫾ 16.4 (60 ml)
42.1 ⫾ 12.7 (60 ml)
32.0 ⫾ 5.7 (120 ml)
12.6 ⫾ 6.0 (60 ml)
4.8 ⫾ 0.8 (60 ml)
4.3 ⫾ 0.8 (120 ml)
a
n ⫽ 10 for each condition.
* Surgical surface at 100 mm Hg significantly different than all group comparisons (P ⬍
0.001).
† Surgical surface at 300 mm Hg significantly different that beaker at 100 mm Hg, beaker at
300 mm Hg, and beaker at 300 mm Hg ⫹ dilution (P ⬍ 0.001).
‡ Surgical surface at 300 mm Hg ⫹ dilution significantly different than beaker at 100 mm
Hg, beaker at 300 mm Hg, and beaker at 300 mm Hg ⫹ dilution (P ⬍ 0.001).
suction from both a beaker and the flat surface by
approximately 60%.
DISCUSSION
Hemolysis ranged from 0.21% to 2.29% in this
study. In the only other study of this topic, Gregoretti
(6) reported hemolysis ranging from 0.32% ⫾ 0.21%
when blood alone was aspirated to 2.85% ⫾ 0.22%
(P ⬍ 0.05) when a blood–air admixture was suctioned
at a vacuum pressure of 300 mm Hg, similar to the
values in our study.
Mathematical modeling of cell salvage has revealed
that small changes in RBC processing efficiency can
make large differences in the maximum allowable
blood loss that a patient can sustain before allogeneic
transfusion therapy (14,15). These models suggest that
a 70 kg patient with a starting HCT of 45% can sustain
a blood loss of 9600 mL if a transfusion trigger of 21%
is used and cell salvage captures 60% of lost RBCs. The
sustainable blood loss increases to 13,750 mL if 70%
RBC recovery is achieved. Thus, small changes in RBC
recovery can result in large differences in the ability to
avoid allogeneic transfusion.
In an attempt to maximize capture of RBCs, many
cell salvage programs will minimize suction pressure
applied to the RBCs when it is being removed from
the surgical field. Decreasing the suction pressure
decreases the shear forces applied to the RBCs, which
in turn decreases hemolysis. In addition, surgical
686
Minimizing Mechanical Blood Damage
sponges are rinsed with saline before discard with the
rinse solution being processed through the cell salvage
cycle, resulting in capture of these RBCs. Regulation of
suction pressure and rinsing of sponges are the only
known techniques for maximizing efficiency of RBC
recovery.
The visual appearance of the blood after high
suction from the flat surface showed heavy amounts
of frothing from aspirated air. Diluting blood with
normal saline reduced the frothing and mechanical
stress, resulting in an approximately 60% reduction in
hemolysis. This suggests that hemolysis can be reduced in some surgical environments by instilling
normal saline into the surgical field during suctioning.
For example, spine surgery involves a small suction
tip on a relatively flat, shallow surgical field. In these
procedures, use of saline along with suctioning would
facilitate RBC recovery. The same effect might be
achievable by combining normovolemic hemodilution
with cell salvage.
The highest physiological shear stress in the normal
vasculature is about 10 N/m2 (16). The highest wall
shear stresses applied to blood in our experiments
were about 20 N/m2. This shear stress is significantly
lower than those of 200 –300 N/m2 considered to be
lethal for RBCs (17). This suggests that other factors
are more important than the aspiration method in the
RBC loss during blood salvage.
The blood–air interface may also contribute to
increased hemolysis during filtration at the point of
the collection, and the concentrating and washing
processes. Sublethal RBC damage is analogous to
“accelerated” RBC aging, and promotes early removal
of the transfused RBCs from the vascular system (18).
Our results confirm the previous work of Gregoretti,
which showed that suction of air should be avoided to
enhance RBC capture and return. The degree of hemolysis observed in both studies does not explain the
large differences in capture rates obtained in Waters’
mathematical modeling (16). The reason for this could
be a sublethal blood damage produced during the first
stage of blood moving from a surgical site to its final
mixing with saline before it is returned to a patient.
Sublethal damage, which is not easy to identify, could
significantly reduce RBC survival during washing and
filtration. Minimizing suction pressure will decrease
ANESTHESIA & ANALGESIA
sublethal damage and increase the efficiency RBC
salvage.
REFERENCES
1. Wallace EL, Churchill WH, Surgenor DM, et al. Collection and
transfusion of blood and blood components in the United States,
1994. Transfusion 1998;38:625–36.
2. Blajchman MA. Transfusion-associated immunomodulation
and universal white cell reduction: are we putting the cart
before the horse? Transfusion 1999;39:665–70.
3. Moore SB. Transfusion-related acute lung injury (TRALI): clinical presentation, treatment, and prognosis. Crit Care Med
2006;34(5 Suppl):S114 –17.
4. Kleinman S, Chan P, Robillard P. Risks associated with transfusion of cellular blood components in Canada. Transfus Med
Rev 2003;17:120 – 62.
5. Waters JH, Lee JS, Karafa MT. A mathematical model of cell
salvage compared and combined with normovolemic hemodilution. Transfusion 2004;44:1412–16.
6. Gregoretti S. Suction-induced hemolysis at various vacuum
pressures: implications for intraoperative blood salvage. Transfusion 1996;36:57– 60.
7. Malinauskas RA, Sade RM, Dearing JP, et al. Blood damaging
effects of cardiotomy suction return. J Extra Corpor Technol
1988;20:40 –5.
Vol. 104, No. 3, March 2007
8. Leverett LB, Hellums JD, Alfrey CP, Lynch EC. Red blood cell
damage by shear stress. Biophys J 1972;12:257–73.
9. Lee WH Jr, Hairston P. Structural effects on blood proteins at
the gas-blood interface. Fed Proc 1971;30:1615–22.
10. Galletti PM. Blood interfacial phenomena: an overview. Fed
Proc 1971;30:1491–3.
11. Pierce EC II. The membrane versus bubble oxygenator controversy. Ann Thorac Surg 1980;29:49 –79.
12. Boonstra PW, Vermeulen FE, Leusink JA, et al. Hematological
advantage of a membrane oxygenator over a bubble oxygenator
in long perfusions. Ann Thorac Surg 1986;41:297–300.
13. Mizuguchi K, Damm GA, Aber GS, et al. Does hematocrit affect in
vitro hemolysis test results? Preliminary study with Baylor/NASA
prototype axial flow pump. Artif Organs 1994;18:650 – 6.
14. Hay SN, Monk TG, Brecher ME. Intraoperative blood salvage: a
mathematical perspective. Transfusion 2002;42:451–5.
15. Waters JH, Lee JS, Karafa M. Mathematical modeling of cell
salvage efficiency. Anesth Analg 2002;95:1312–7.
16. Lowe GDO, ed. Clinical blood rheology. Boca Raton, FL: CRC
Press, 1988.
17. Nevaril CG, Lynch EC, Alfrey CP, Hellums JD. Erythrocyte
damage and destruction induced by shearing stress. J Lab Clin
Med 1968;71:784 –90.
18. Kameneva MV, Antaki JF, Borovetz HS, et al. Mechanisms of
red blood cell trauma in assisted circulation. Rheologic similarities of red blood cell transformations due to natural aging and
mechanical stress. ASAIO J 1995;41:M457– 60.
© 2007 International Anesthesia Research Society
687