Photoplethysmography and its application in clinical physiological measurement
- PMID: 17322588
- DOI: 10.1088/0967-3334/28/3/R01
Photoplethysmography and its application in clinical physiological measurement
Abstract
Photoplethysmography (PPG) is a simple and low-cost optical technique that can be used to detect blood volume changes in the microvascular bed of tissue. It is often used non-invasively to make measurements at the skin surface. The PPG waveform comprises a pulsatile ('AC') physiological waveform attributed to cardiac synchronous changes in the blood volume with each heart beat, and is superimposed on a slowly varying ('DC') baseline with various lower frequency components attributed to respiration, sympathetic nervous system activity and thermoregulation. Although the origins of the components of the PPG signal are not fully understood, it is generally accepted that they can provide valuable information about the cardiovascular system. There has been a resurgence of interest in the technique in recent years, driven by the demand for low cost, simple and portable technology for the primary care and community based clinical settings, the wide availability of low cost and small semiconductor components, and the advancement of computer-based pulse wave analysis techniques. The PPG technology has been used in a wide range of commercially available medical devices for measuring oxygen saturation, blood pressure and cardiac output, assessing autonomic function and also detecting peripheral vascular disease. The introductory sections of the topical review describe the basic principle of operation and interaction of light with tissue, early and recent history of PPG, instrumentation, measurement protocol, and pulse wave analysis. The review then focuses on the applications of PPG in clinical physiological measurements, including clinical physiological monitoring, vascular assessment and autonomic function.
Similar articles
-
Photoplethysmography and nociception.Acta Anaesthesiol Scand. 2009 Sep;53(8):975-85. doi: 10.1111/j.1399-6576.2009.02026.x. Epub 2009 Jun 30. Acta Anaesthesiol Scand. 2009. PMID: 19572939
-
[Principles of photoplethysmography and its applications in physiological measurements].Sheng Wu Yi Xue Gong Cheng Xue Za Zhi. 2013 Aug;30(4):899-904. Sheng Wu Yi Xue Gong Cheng Xue Za Zhi. 2013. PMID: 24059078 Review. Chinese.
-
Respiration can be monitored by photoplethysmography with high sensitivity and specificity regardless of anaesthesia and ventilatory mode.Acta Anaesthesiol Scand. 2005 Sep;49(8):1157-62. doi: 10.1111/j.1399-6576.2005.00721.x. Acta Anaesthesiol Scand. 2005. PMID: 16095458 Clinical Trial.
-
Combined photoplethysmographic monitoring of respiration rate and pulse: a comparison between different measurement sites in spontaneously breathing subjects.Acta Anaesthesiol Scand. 2007 Oct;51(9):1250-7. doi: 10.1111/j.1399-6576.2007.01375.x. Epub 2007 Aug 15. Acta Anaesthesiol Scand. 2007. PMID: 17711563
-
Noninvasive monitoring by photoplethysmography.Clin Perinatol. 2012 Sep;39(3):573-83. doi: 10.1016/j.clp.2012.06.012. Clin Perinatol. 2012. PMID: 22954270 Review.
Cited by
-
Heart Rate Variability Measurement through a Smart Wearable Device: Another Breakthrough for Personal Health Monitoring?Int J Environ Res Public Health. 2023 Dec 6;20(24):7146. doi: 10.3390/ijerph20247146. Int J Environ Res Public Health. 2023. PMID: 38131698 Free PMC article. Review.
-
Innovative continuous non-invasive cuffless blood pressure monitoring based on photoplethysmography technology.Intensive Care Med. 2013 Sep;39(9):1618-25. doi: 10.1007/s00134-013-2964-2. Epub 2013 Jun 6. Intensive Care Med. 2013. PMID: 23740275
-
Improved Hypertension Risk Assessment with Photoplethysmographic Recordings Combining Deep Learning and Calibration.Bioengineering (Basel). 2023 Dec 18;10(12):1439. doi: 10.3390/bioengineering10121439. Bioengineering (Basel). 2023. PMID: 38136030 Free PMC article.
-
Energy-Efficient PPG-Based Respiratory Rate Estimation Using Spiking Neural Networks.Sensors (Basel). 2024 Jun 19;24(12):3980. doi: 10.3390/s24123980. Sensors (Basel). 2024. PMID: 38931763 Free PMC article.
-
Classification of Diabetes Using Photoplethysmogram (PPG) Waveform Analysis: Logistic Regression Modeling.Biomed Res Int. 2020 Aug 11;2020:3764653. doi: 10.1155/2020/3764653. eCollection 2020. Biomed Res Int. 2020. Retraction in: Biomed Res Int. 2024 Mar 20;2024:9808737. doi: 10.1155/2024/9808737. PMID: 32851065 Free PMC article. Retracted.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources