Abstract
Purpose
To measure the extent of corneal displacement during the early phase of tonometry using a noncontact tonometer and to determine the factors that affect the extent of the displacement.
Methods
The cornea was photographed in profile by use of a high-speed camera during tonometry with a noncontact tonometer. The frame rate of the camera was 5,000 frames/s. The extent of the displacement of the central corneal area at 13.2 ms after application of the air puff was measured in 115 healthy volunteers. The factors that affected the extent of the corneal displacement were determined by stepwise multiple linear regression analysis.
Results
The subjects’ age, sex, intraocular pressure (IOP), central corneal thickness, anterior chamber depth, and axial length were selected by the stepwise method. The results of the multiple linear regression analysis showed that subjects with low IOP, of older age, and of male sex had significantly greater displacement of the central cornea.
Conclusions
High-speed photography can be used to evaluate the degree of corneal displacement during tonometry with a noncontact tonometer. The amount of corneal displacement is affected by the individual’s IOP, age, and sex.
Similar content being viewed by others
References
American Academy of Ophthalmology Glaucoma Panel. Preferred practice pattern guidelines. Primary open-angle glaucoma. San Francisco: American Academy of Ophthalmology; 2010. http://www.aao.org/ppp.
Herndon LW, Choudhri SA, Cox T, Damji KF, Shields MB, Allingham RR. Central corneal thickness in normal, glaucomatous, and ocular hypertensive eyes. Arch Ophthalmol. 1997;115:1137–41.
Paranhos A Jr, Paranhos FR, Prata JA, Omi CA, Mello PA, Shields MB. Influence of keratometric readings on comparative intraocular pressure measurements with Goldmann, Tono-Pen, and noncontact tonometers. J Glaucoma. 2000;9:219–23.
Doughty MJ, Zaman ML. Human corneal thickness and its impact on intraocular pressure measures: a review and meta-analysis approach. Surv Ophthalmol. 2000;44:367–408.
Damji KF, Muni RH, Munger RM. Influence of corneal variables on accuracy of intraocular pressure measurement. J Glaucoma. 2003;12:69–80.
Emara B, Probst LE, Tingey DP, Kennedy DW, Willms LJ, Machat J. Correlation of intraocular pressure and central corneal thickness in normal myopic eyes and after laser in situ keratomileusis. J Cataract Refract Surg. 1998;24:1320–5.
Friedenwald JS. Contribution to the theory and practice of tonometry. Am J Ophthalmol. 1937;20:985–1024.
Pallikaris IG, Kymionis GD, Ginis HS, Kounis GA, Tsilimbaris MK. Ocular rigidity in living human eyes. Invest Ophthalmol Vis Sci. 2005;46:409–14.
Elsheikh A, Wang D, Pye D. Determination of the modulus of elasticity of the human cornea. J Refract Surg. 2007;23:808–18.
Knox Cartwright NE, Tyrer JR, Marshall J. Age-related differences in the elasticity of the human cornea. Invest Ophthalmol Vis Sci. 2011;52:4324–9.
Kempf R, Kurita Y, Iida Y, Kaneko M, Mishima H, Tsukamoto H, et al. Dynamic properties of human eyes. Conf Proc IEEE Eng Med Biol Soc. 2005;3:3180–3.
Kempf R, Kurita Y, Iida Y, Kaneko M, Sugimoto E, Tsukamoto H, et al. Understanding eye deformation in non-contact tonometry. Conf Proc IEEE Eng Med Biol Soc. 2006;1:5428–31.
Newsome DA, Foidart JM, Hassell JR, Krachmer JH, Rodrigues MM, Katz SI. Detection of specific collagen types in normal and keratoconus corneas. Invest Ophthalmol Vis Sci. 1981;20:738–50.
Ricard-Blum S, Ville G. Collagen cross-linking. Int J Biochem. 1989;21:1185–9.
Richards SC, Olson RJ, Brodstein DE, Richards WL. Differences between men and women as related to intraocular lens implantation. Ophthalmic Surg. 1986;17:82–7.
Goto T, Klyce SD, Zheng X, Maeda N, Kuroda T, Ide C. Gender- and age-related differences in corneal topography. Cornea. 2001;20:270–6.
Weinreb RN, Lu A, Beeson C. Maternal corneal thickness during pregnancy. Am J Ophthalmol. 1988;105:258–60.
Wickham LA, Rocha EM, Gao J, Krenzer KL, da Silveira LA, Toda I, et al. Identification and hormonal control of sex steroid receptors in the eye. Adv Exp Med Biol. 1998;438:95–100.
Luce DA. Determining in vivo biomechanical properties of the cornea with an ocular response analyzer. J Cataract Refract Surg. 2005;31:156–62.
Kotecha A, Elsheikh A, Roberts CR, Zhu H, Garway-Heath DF. Corneal thickness- and age-related biomechanical properties of the cornea measured with the ocular response analyzer. Invest Ophthalmol Vis Sci. 2006;47:5337–47.
Allingham RR. Shields’ textbook of glaucoma. 5th ed. Philadelphia: Lippincott Williams & Wilkins; 2005. p. 36–58.
Acknowledgment
This work was partially supported by grant no. 22591965 from the Japanese Ministry of Education, Culture, Science and Technology.
Author information
Authors and Affiliations
Corresponding author
About this article
Cite this article
Kiuchi, Y., Kaneko, M., Mochizuki, H. et al. Corneal displacement during tonometry with a noncontact tonometer. Jpn J Ophthalmol 56, 273–279 (2012). https://doi.org/10.1007/s10384-012-0131-x
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s10384-012-0131-x